Place of Origin:CN;GUA Brand Name:Drager Model Number:OOM 201 OOM 201 Drager oxygen sensor
Item #: B226-205Manufacturer: Westprime HealthcareManufacturer's Product Number: OOM201 / E61663Condition: New
Sensor Specifications & Compatibility
Catalog Number
E61663 /OOM201
Output Signal in Air
24A - 32A (Dual Cathode) no temperature compensation
Response Time T 90%
< 12 seconds
Nominal Sensor Lifetime
500 000 % volume Oxygen hours
Electrical Interface
Gold plated slip rings
Compatibility:
Draeger 6850645 (Dr¤ger)
Babylog, Fabian, Cicero, PM8030
Evita, Oxydig, Anemone, Oxycom
Incubator 8000
M08 ST13
MAX11
Invivo Magnitude
10-103-07
PT14A
Teledyne R13 C-61139
Hudson 5557 (V14)
VTI 101372, 100352
Medical Oxygen Sensors (Fuel Cells) general specifications:
Measurement Range:
0-100 % oxygen
Accuracy and Repeatability:
< 1 % vol. O 2 when calibrated at 100 % O 2
Zero Ofset:
< 0.5 % vol. O 2 in 100% N 2 , applied 5 minutes
Linearity Error:
< 3 % relative
Cross Interference:
< 0.5 % vol. O 2 response to:
10 % CO 2 balance N 2
80% N 2 O balance N 2
7.5% Halothane balance N 2
7.5 % Isofurane balance N 2
7.5 % Enfurane balance N 2
9% Sevofurane balance N 2
20% Desfurane balance N 2
Infuence of Humidity:
- 0.03 % rel. per % RH at 25C
Infuence of Pressure:
proportional to change in oxygen partial
pressure
Infuence of Mechanical Shock:
< 1% relative after a fall from 1m
Operating Temperature:
0 to 50C
Temperature Compensation:
built-in NTC compensation (see below)
Efect of Temperature
between +5 C and +25 C: 3 % relative error
Compensation (steady state):
between 0 C and +50 C: 8 % relative error
Operating Humidity:
0-99 % RH non-condensing
Long Term Output Drift:
< 1 % vol oxygen per month
typically < - 15 % relative over lifetime
Storage Temperature:
-20 to +50 C
Recommended Storage:
+5 to +15 C
Recommended Load:
10 kOhms
Warm-Up Time:
< 30 minutes, after replacement of sensor
Weight:
approximately 28 grams
Certificates
CMEF & ARAB Exhibition
Upnmed Home
Upnmed Factory
Testing
Terms of Payment
Type of Shipping
Related Products
Model NO. Compatible HP temperature sensor
Model NO. MR850 Respiratory Humdifier
Wechat scan code or copy link mobile phone open
X
OOM 201 Drager oxygen sensor
Send
OOM 201 Drager oxygen sensor
Send
13670105485
This page requires the browser to support (enable) JavaScript