UpnMed Co.,Ltd518106About UsNo.301,Building #41st ,The 3rd Industrial Zone, Xitian Community, Gongming Street, Guangming District, Shenzhen,China,Zip:518106UpnMed is a leading professional manufacturer specializing in Medical Accessories, with more than 10 years experiences in this field.  Our production line covers: Reusable Spo2 Sensor, Disposable Spo2 Sensor,Spo2 Extension Cable, ECG cable, EKG cable ,IBP cable and Disposable pressure transducer, NIBP cuff & tube, Temperature Probe, Pulse Oximeter, B ultrasound probe, ESU accessories, Fetal & TOCO probe, OEM & ODM services etc.   We have both legal production certifications and accuracy test equipment. Most of our products are verified by CFDA,CE,Free Sales Certification and

OOM 201 Drager oxygen sensor
OOM 201 Drager oxygen sensor

OOM 201 Drager oxygen sensor

Model NO.
OOM 201
Price:
US $ 75.00-75.00 / piece
Minimum order quantity
1 piece
Supply Ability
1000 piece / Week
Country of Origin
Shenzhen
Quantity:
  • Place of Origin:CN;GUA
  • Brand Name:Drager
  • Model Number:OOM 201

 OOM 201 Drager oxygen sensor

 

  • Item #: B226-205
  • Manufacturer: Westprime Healthcare
  • Manufacturer's Product Number: OOM201 / E61663
  • Condition: New

 

Sensor Specifications & Compatibility

 

Catalog Number

E61663/OOM201

Output Signal in Air

24A - 32A (Dual Cathode) no temperature compensation

Response Time T 90%

< 12 seconds

Nominal Sensor Lifetime

 500 000 % volume Oxygen hours

Electrical Interface

Gold plated slip rings

Compatibility:

 

 

 

 

 

 

 

 

 

 

 

Draeger 6850645 (Dr¤ger)

Babylog, Fabian, Cicero, PM8030

Evita, Oxydig, Anemone, Oxycom

Incubator 8000

M08 ST13

MAX11

Invivo Magnitude

10-103-07

PT14A

Teledyne R13 C-61139

Hudson 5557 (V14)

VTI 101372, 100352

 

Medical Oxygen Sensors (Fuel Cells) general specifications:

Measurement Range:

0-100 % oxygen

Accuracy and Repeatability:

< 1 % vol. O2when calibrated at 100 % O2

Zero Ofset:

< 0.5 % vol. O2in 100% N2, applied 5 minutes

Linearity Error:

< 3 % relative

Cross Interference:

 

 

 

 

 

 

 

< 0.5 % vol. O2response to:

10 % CO2balance N2

80% N2O balance N2

7.5% Halothane balance N2

7.5 % Isofurane balance N2

7.5 % Enfurane balance N2

9% Sevofurane balance N2

20% Desfurane balance N2

Infuence of Humidity:

- 0.03 % rel. per % RH at 25C

Infuence of Pressure:

proportional to change in oxygen partial

pressure

 

Infuence of Mechanical Shock:

< 1% relative after a fall from 1m

Operating Temperature:

0 to 50C

Temperature Compensation:

built-in NTC compensation (see below)

Efect of Temperature

between +5 C and +25 C: 3 % relative error

Compensation (steady state):

between 0 C and +50 C: 8 % relative error

Operating Humidity:

0-99 % RH non-condensing

Long Term Output Drift:

< 1 % vol oxygen per month

typically < - 15 % relative over lifetime

 

Storage Temperature:

-20 to +50 C

Recommended Storage:

+5 to +15 C

Recommended Load:

10 kOhms

Warm-Up Time:

< 30 minutes, after replacement of sensor

Weight:

approximately 28 grams

 

 

 

 

 

 

 

 

Certificates

 

 

CMEF & ARAB Exhibition

 

 